System Performance Information
System Performance Information
Precision and Accuracy Summary:
Accuracy: The accuracy of the CONTOUR
®
USB blood
glucose monitoring system has been assessed in a
correlation study conducted by healthcare professionals. In
this study 100 fresh capillary blood specimens were tested
with the CONTOUR USB blood glucose monitoring
system by 100 untrained people with diabetes using three
test strip lots. The reference measurement was done on
the YSI™ Glucose analyzer. CONTOUR USB data compare
well with the results from the YSI method.
Regression Statistics (Plasma/Serum Reference):
plasma reference glucose range: 3.2 mmol/L to 22.8 mmol/L
Lot y =
Correlation
Coeffi cient (r)
1 0.97x + 0.25 mmol/L 0.975
2 1.04x – 0.23 mmol/L 0.971
3 1.01x + 0.15 mmol/L 0.979
Precision: A study was conducted with the CONTOUR
USB blood glucose monitoring system with fi ve heparinized
whole blood specimens with glucose levels from 2.3 mmol/L
to 18.9 mmol/L. Multiple replicates (n=100) were tested
using multiple CONTOUR
®
USB blood glucose meters and
one lot of CONTOUR
®
blood glucose test strips. The
following precision results were obtained.
Plasma/Serum:
Mean SD
Coeffi cient
of variation
2.4 mmol/L 0.06 mmol/L 2.4%
4.7 mmol/L 0.08 mmol/L 1.7%
7.3 mmol/L 0.10 mmol/L 1.3%
11.3 mmol/L 0.19 mmol/L 1.7%
18.4 mmol/L 0.21 mmol/L 1.2%
Limitations:
Preservatives: Blood may be collected by healthcare
professionals into test tubes containing heparin.
Do not use other anticoagulants or preservatives.
Altitude: Up to 3048 meters (10,000 feet) does not
signifi cantly affect results.
Lipemic Specimen: Cholesterol concentrations
>12.9 mmol/L or triglyceride concentrations
>34.3 mmol/L may produce elevated readings.
Peritoneal dialysis solutions: Icodextrin does not
interfere with CONTOUR
®
test strips.
Xylose: Do not use during or soon after xylose
absorption testing. Xylose in the blood will cause
an interference.
Contraindications: Capillary blood glucose testing
may not be clinically appropriate for persons with
reduced peripheral blood fl ow. Shock, severe
hypotension, hyperosmolar hyperglycemia and severe
dehydration are examples of clinical conditions that
may adversely affect the measurement of glucose in
peripheral blood.
4
Interference: Reducing substances occurring in the
blood naturally (uric acid, bilirubin) or from therapeutic
treatments (ascorbic acid, acetaminophen) will not
signifi cantly affect results. The limiting concentrations
of these compounds are listed in your test strip insert.
System Requirements:
Windows
®
:
XP SP3, Vista SP2
Mac: OS
®
X v10.5.2 (Intel
®
64 bit processors only)
High-Power USB port
Java 1.6.0_07 or higher
For latest compatibility: www.bayerdiabetes.ca